|
Catalog Number RBY4C0415-A |
Device Problem
Component Falling (1105)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/27/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the pet lock was intact on the proximal end of the ruby coil pusher assembly; the coil was intact with the pusher assembly; the introducer sheath was not returned for evaluation.Conclusions: evaluation of the returned device revealed no visible damage to the ruby coil.The ruby coil's introducer sheath was not returned for evaluation.Ruby coils are 100% functionally tested during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the physician deployed and detached six ruby coils into the aneurysm using a lantern delivery microcatheter (lantern).While a new ruby coil was being loaded into the lantern, the ruby coil unintentionally fell off the table.The dropped ruby coil was not used for the procedure and did not enter the patient's body.The procedure was completed using two new ruby coils.
|
|
Search Alerts/Recalls
|
|
|