Catalog Number RBY2C0435-A
Device Problems Detachment Of Device Component (1104); Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2016
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2016-00248.Device was disposed of by hospital.
 
Event Description
The patient was undergoing a coil embolization procedure using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, while advancing a ruby coil through the lantern, the physician met resistance and decided to resheath the ruby coil.While the ruby coil was being retracted, it unintentionally detached inside the lantern and became stuck.The physician was unable to flush the coil with saline or retract the coil back out and therefore, decided to remove the lantern containing the ruby coil from the patient.The procedure was completed using two new ruby coils and another manufacturer's microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
kathleen kidd
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key5464583
MDR Text Key39520740
Report Number3005168196-2016-00247
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548012889
UDI-Public00814548012889
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/09/2023
Device Catalogue NumberRBY2C0435-A
Device Lot NumberF65635
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
-
-