Method: the complaint rd900 infant resuscitator was returned to our service center in irvine, california, where it was visually inspected and performance tested.Results: visual inspection revealed cracks on the enclosure on the upper and lower end caps.Cracks were also noted on the fascia and valve assembly.Performance test revealed that the manometer was faulty.A lot check revealed no other complaints of this nature for lot 070816.Conclusion: the neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.It must be noted that the subject neopuff is over eight years old.All neopuff units are visually inspected and performance tested prior to leaving the production line, and those that fail are rejected.The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly.If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested.The neopuff technical manual states the following: "dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient".
|