• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES BRANDYWINE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE,DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES BRANDYWINE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE,DEPTH Back to Search Results
Catalog Number 319.006
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2016
Event Type  malfunction  
Manufacturer Narrative
This device was used for treatment, not diagnosis.(b)(4).Device is an instrument and is not implanted/explanted.Subject device has been received and is currently in the evaluation process.No service history review can be performed as this is a lot controlled item.The manufacture date of this item is 20-may-2004.The source of the manufacture date is the release to warehouse date.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported the service and repair department documented that the needle broke off while going through the wash.There was no case or patient involvement.There is no additional information.This report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Synthes manufacturing location was corrected upon findings.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.A service and repair evaluation was performed for the subject device (depth gauge, part number 319.006, lot number 4772131).The customer reported the needle broke off.The repair technician reported ¿tip broken¿ as the reason for repair.The item is not repairable.The cause of the issue is unknown.The service and repair evaluation was confirmed.The subject device was forwarded to synthes customer quality for additional investigation.A product development investigation was performed for the subject device by synthes customer quality.The associated product drawing was reviewed as part of the investigation.No drawing issues or discrepancies were noted.The design is adequate for its intended use and did not contribute to this complaint condition.The thickness of the needle (1.25 mm) is driven by the fact that the needle must fit into a drilled hole of 1.5 mm, and the length (80 mm) is determined so the slider can measure screws up to 40 mm.The material of the needle probe component (part # 319.006.03) is extra hard 316ss, which is an appropriate material for an instrument component of this type.A device history record (dhr) review was performed for the subject device lot.The review of the dhrs for the top-level and each subcomponent showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No non-conformances were generated during the production of the subject device.The returned depth gauge shows regular use during its lifespan.The hooked needle stem of the device is broken off at the base of the black body (the broken stem is approx.75mm in length) and was not returned.This particular depth gauge is part of at least 14 technique guides, including the 2.4 mm variable angle lcp distal radius system and is used to measure the depth of the holes for the 2.0mm/2.4mm screws to ensure the correct screw length is used during the procedure.The information is provided per the 2.4 mm variable angle lcp distal radius system technique guide.Although the damage appears to be the result of excessive weight being placed onto the needle during sterile processing and does not appear to be the result of normal use; the exact cause could not be identified.This complaint is confirmed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE,DEPTH
Manufacturer (Section D)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer (Section G)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5480787
MDR Text Key39930737
Report Number1719045-2016-10192
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number4772131
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-