The patient was implanted with a left ventricular assist device.During a clinic visit, a cut in the outer coating of the driveline was observed that the patient reported had occurred the week prior.Upon manipulation of the area, the system controller sounded low speed advisories and a pump stoppage occurred.The patient was asymptomatic.On (b)(6) 2016, the patient's driveline was repaired by the manufacturer's technical services representatives.It was reported that the repair appeared to be successful.No subsequent events have been reported.
|
Device evaluation: the evaluation of the returned portion of the driveline confirmed an issue that would have contributed to the reported low speed events and pump stoppage.A distal end driveline replacement was performed and approximately 15 inches of the external portion/distal end of the driveline was returned for evaluation.Electrical continuity testing of the driveline segment revealed that all wires were electrically intact.The tongue depressors, cotton, and tape which were previously used to stabilize the driveline were removed, and the reported area of superficial damage to the outer silicone sleeve was noted at approximately 4.75 inches from the metal connector.Upon removal of the outer silicone sleeve, the clear bionate layer appeared unremarkable.Breakdown of the metal braided shield was observed near the terminus of the distal end bend relief at approximately 2.5-3.5 inches from the metal connector.Visual inspection of the underlying wires in the location of shield breakdown found that the orange wire was partially fractured at approximately 2.75 inches from the metal connector.The observed wire damage appeared to be the result of fatigue failure due to repetitive flexing of the driveline in this area.If the exposed conductors of the orange wire made contact with the braided shield while the patient was operating on a tethered power source such as the power module, the resulting electrical short to ground would have caused the reported low speed events and pump stoppage that were confirmed via the submitted system controller log file.A review of the device history records revealed the device met applicable specifications.No further information was provided.The manufacturer is closing the file on this event.
|