Catalog Number 8065990601 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.Additional information has been requested.(b)(4).
|
|
Event Description
|
A customer reported erratic readings with the micrometer.Upon follow up, reporter indicated the micrometer erratic measurements occurred during the calibration before surgery.They were able to get the micrometer to stabilize, after speaking with the field service engineer over the phone.Reporter relayed they were able to perform surgery on the event day without any impact on surgery.However; once surgery day was complete, and the micrometer was tested again, they again experienced erratic measurements.
|
|
Manufacturer Narrative
|
Sample has not been returned for evaluation.During on site servicing visit the field service engineer (fse) replaced micrometer and verified the system.The most likely root cause was a faulty micrometer.Deeper root cause investigation is not possible without sample.(b)(4).
|
|
Search Alerts/Recalls
|