Model Number 1863
Device Problem Invalid Sensing (2293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/10/2016
Event Type  malfunction  
Manufacturer Narrative
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
 
Event Description
It was reported that the spo2 measurement value showed 80's%.The patient was not so bad health condition, so it must be the high 90's%.Also, the display didn't show normal waveform.No known impact or consequence to patient.
 
Manufacturer Narrative
The returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.The sensor was determined to be functioning as designed.A review of the lot information was performed and there were no previous reported issues prior to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LNCS DC-I
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO CORPORATION
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001,
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
charlene johnson
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key5488191
MDR Text Key39934705
Report Number2031172-2016-00312
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K051212
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1863
Device Catalogue Number1863
Device Lot Number13KBW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-