During preparation for a thrombectomy procedure, the hospital staff noticed that when the penumbra system aspiration pump max 220v (pump max) was turned on, only the fan made a faint noise; no vacuum was produced and there was no pressure reading.The defective pump max was found prior to use and was not used during the procedure.The procedure was completed using a stent retriever.
|
There was no visible damage to the exterior of the pump.The pump was plugged in and powered on.The vacuum pump did not start; only the pump cooling fan started.The pump was opened by penumbra engineers and corrosion was observed on the piston crown in the outlet cylinder.Evaluation of the returned pump confirmed that even though power was supplied to the pump, the vacuum pump would not turn on.The pump housing was removed, and the vacuum pump was opened by penumbra engineers.It was observed that the piston crown in the outlet cylinder was corroded.The observed corrosion likely caused the piston to seize inside the cylinder.The root cause of this corrosion could not be determined.Penumbra pumps are 100% functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|