This device referenced in this report was returned to olympus (b)(4) for evaluation.The bending rubber was torn and it was stretched and folded towards the distal side.The bending section of the subject device worked within the specifications.The surgeon thought that the fragments of the stone came to the proximal side of the bending section and when the scope was withdrawn, the scope became trapped on the undulations, and then the bending section of the scope was damaged.The subject device became trapped possibly by stones between the subject device and the ureter.The manufacturing record of the subject device was reviewed without irregularity.If additional information becomes available at a later time, this report will be supplemented.
|
The device referenced in this report was sent to (b)(4) olympus for evaluation.According to the evaluation, the bending rubber was folded on the distal side and there was a cutting found on the proximal side.In addition, it was confirmed that there was a sliding mark on the insertion tube, and a cutting, uncounted pinholes on the bending rubber between 90mm and 50mm from the distal end on the up side, and between 90mm and 30mm on the down side.Nothing abnormal was confirmed on the bending shape and the bending movement.Further, a stain was found on the imager guide.From the above analysis, as a cause of this event, it is considered that the bending rubber was folded on the distal side and the diameter became larger, causing the scope to be stuck in the ureter.Though it has not been informed whether the user facility used the access sheath or not, if the access sheath was used, there is a possibility the distal end of the bending rubber was folded because the surgeon forcefully tried to withdraw the scope that was stuck in the access sheath.For the cause of the stain on the imager guide, it is considered that moisture was flooded through the cuttings or pinholes of the bending rubber.
|