The patient was implanted with a left ventricular assist device (lvad).It was reported that on (b)(6) 2016, while prepping and draping the patient for a heart transplant, a sudden pump stoppage occurred and the system controller alarmed ''connect driveline.'' at the time of the pump stoppage, no one was touching the patient, system controller, percutaneous lead, patient cable or the power module.It was reported that the percutaneous lead was fully engaged.The hospital staff disconnected the percutaneous lead and reinserted it, without resolution to the alarm.The patient was placed on battery power in case of a short to shield condition but the alarm did not resolve.It was reported that patient became hemodynamically unstable and the hospital staff exchanged the system controller with the patient's backup system controller and the pump started.The patient was prepped and draped, and right after the surgical incision was made pump stoppages re-occurred.The percutaneous lead was jiggled, and the pump started and stopped.The hospital staff was able to maintain pump function by holding the percutaneous lead in just the right position for 1.5 hours, until the patient safely went on bypass.The event could not be reproduced again.The patient was transplanted and no further issues were reported.
|
The evaluation of the returned pump confirmed internal driveline wire damage that could have contributed to the reported event.The pump was returned assembled with the driveline cut approximately 2.5 inches from the pump housing and the remainder of the driveline was returned in two segments measuring approximately 6 inches and 30.5 inches long, respectively.Visual examination the driveline segments revealed a breach in the insulation of the brown wire, approximately 2 inches from one of the cut ends of the 6 inches long segment.High potential testing revealed no additional areas of current leakage.If the exposed conductors of the brown wire contacted with the braided shield while the system was operating on the power module, the resulting short to ground would have interrupted pump function and resulted in the reported red heart alarms and pump stoppages.A review of the device history records revealed no deviations from manufacturing or quality assurance specifications.The manufacturer is closing the file on this event.
|