MAUDE Adverse Event Report: MAKO SURGICAL CORP. TIBIAL INSERT ONLAY TRIAL-SIZE 2-8MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. TIBIAL INSERT ONLAY TRIAL-SIZE 2-8MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 170702-1
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/18/2016
Event Type  malfunction  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was discarded and was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event Description
A 2x8 poly was damaged during trialing.
 
Manufacturer Narrative
Reported event: an event regarding damage involving a tibial insert onlay trial-size 2-8mm was reported.The event was not confirmed.Method & results: device evaluation and results: could not be performed as no items associated with the event were returned or made available for identification or evaluation.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for this lot.Conclusions: the event could not be confirmed nor the root cause of the reported event determined due to the minimal information received for evaluation.No further investigation for this event is possible at this time as insufficient information was received by stryker orthopaedics.If additional information becomes available, this investigation will be reopened.
 
Event Description
A 2x8 poly was damaged during trialing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INSERT ONLAY TRIAL-SIZE 2-8MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
sandra spokane
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5500029
MDR Text Key40327965
Report Number3005985723-2016-00083
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number170702-1
Device Lot Number12060809
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/21/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-