• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA COMPANION HOSPITAL CART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA COMPANION HOSPITAL CART Back to Search Results
Catalog Number 397003-001
Device Problem Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The companion hospital cart was not in use by a patient.The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The hospital cart can be plugged into external wall power to provide a redundant power source to the docked companion 2 driver.The syncardia clinical support specialist reported that the companion hospital cart power cord "falls out" when the hospital cart is being repositioned.The customer, cleveland clinic, has to use tape to hold the power cord in place.This alleged failure mode poses a low risk to a patient because it would not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver has redundant, alternate power sources of external batteries and an internal, emergency battery.Syncardia initiated a capa (corrective and preventive action) to address the companion driver caddy power cord issue.The root cause investigation is ongoing.Potential corrective actions will be evaluated when the root cause investigation has been completed.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION HOSPITAL CART
Type of Device
HOSPITAL CART
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5500887
MDR Text Key40835392
Report Number3003761017-2016-00089
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number397003-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/22/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-