A customer in (b)(6) reported errors 6218-ft-026 on their thinprep 5000 processor.It was reported that the instrument sucked up all the liquid in the filter and released it again.Two samples were lost, patient recall is needed.Instrument is not operational, hologic field service engineer dispatched.Fse confirmed and reproduced the error.Found the following to be the most likely cause of the error, filter rotator motor.Replaced the following parts per technical documentation to resolve the error.In addition the rotary vane pump was very slow.Replaced the following parts as precautionary measure.Ran go no-go test to verify instrument operation.Ran sample processing test to verify instrument operation instrument operational.Although the instrument produced an error code during this incident, this is a reportable event since the patient needed to be recalled for additional sample collection, which resulted in a delay in patient diagnosis.
|