• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SCD 700 COMPRESSION SYSTEM-US; SCD CONTROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SCD 700 COMPRESSION SYSTEM-US; SCD CONTROLLER Back to Search Results
Model Number 29525
Device Problems Break (1069); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 03/16/2016.An investigation is currently underway.Upon completion, a full detailed investigation will be provided.
 
Event Description
On (b)(6) 2016 the customer initiated a service repair request for an unknown issue.Upon triage on (b)(6) 2016 the service tech found the unit had a cut power cord with burns/thermal damage (with exposed copper wire).
 
Manufacturer Narrative
A review of the information in the complaint file indicates this investigation was performed by a medtronic technical center for the reported condition of; an issue unknown by the customer.During triage the power cord was found to have a cut with burns.Therefore, this report will be based on information provided by the technical center.The scd 700 was evaluated and the customer reported issue was confirmed; the power cord had visible damage with an internal burn.The cause of the reported condition for the damaged power cord was due to customer misuse.The power cord was removed and replaced to correct the reported issue.The unit passed all initial testing after failure analysis repair instructions were completed.Scd 700 was manufactured in 2013.A review of the device history record shows this device was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.Based on the investigation, no corrective action is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCD 700 COMPRESSION SYSTEM-US
Type of Device
SCD CONTROLLER
Manufacturer (Section D)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH 
Manufacturer (Section G)
COVIDIEN
building #10, 789 puxing road
shanghai
CH  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5503770
MDR Text Key40421082
Report Number3006451981-2016-00156
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number29525
Device Catalogue Number29525
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-