Catalog Number 2C8931
Device Problems Failure to Prime (1492); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 02/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Simulated use testing was performed by spiking the device into an in-house solution bag and priming.The device was found to prime and flow with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink multi-port manifold would not prime.This was observed during priming.The tubing was found to be pinched just after the stopcock.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK MULTI-PORT MANIFOLD
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE - HAINA, SAN CRISTOBAL
parque industrial itabo, zona
haina
DR 
Manufacturer (Section G)
BAXTER HEALTHCARE - HAINA, SAN CRISTOBAL
parque industrial itabo, zona
haina
DR  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5527474
MDR Text Key41220737
Report Number1416980-2016-06030
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K932512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date04/13/2020
Device Catalogue Number2C8931
Device Lot NumberDR15D13024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2016
Date Manufacturer Received03/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-