Customer reported that the radical-7 touchscreen does not audibly alarm when the spo2 alarm limits are exceeded for either high or low settings.However, there is a visual alarm (flashing red bell on the docking station and flashing red highlight behind the spo2 value), but there is no visual error message indicating low or high spo2.The customer confirmed that the speaker was working.The speaker will audibly alarm for initial "sensor off" (and shows the corresponding error message), low and high pulse rate (and shows the corresponding error message), and will also sound during pulse tones and power on.The speaker will also variably sound as adjustments are made to alarm and pulse tone volume settings.No known impact or consequence to patient.
|