Model Number SC-1140 |
Device Problem
Telemetry Discrepancy (1629)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
A report was received that during patient's lead revision procedure, the remote control would not connect with the ipg.The physician confirmed that he may have hit the ipg with the electrocautery.The ipg was replaced and the patient was doing well postoperatively.Monopolar electrocautery is a known source of high voltage transient signal and current company labeling warns against the use of monopolar electrocautery.(physician's implant manual 9055940-001).
|
|
Manufacturer Narrative
|
Sc-1140 sn: (b)(4).Device evaluation indicated that the device passed all tests performed.The complaint was not verified.The ipg was linked with a test novi remote control without any issue.The device functional test was performed to ensure the device integrity.No anomalies were found.
|
|
Event Description
|
A report was received that during patient's lead revision procedure, the remote control would not connect with the ipg.The physician confirmed that he may have hit the ipg with the electrocautery.The ipg was replaced and the patient was doing well postoperatively.Monopolar electrocautery is a known source of high voltage transient signal and current company labeling warns against the use of monopolar electrocautery.(physician's implant manual 9055940-001).
|
|
Search Alerts/Recalls
|