Age at time of event: 18 years or older (b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
It was reported that catheter got stuck in lesion.Vascular access was obtained with ipsilateral retrograde approach.The 99% stenosed target lesion was located in the moderately tortuous and moderately calcified external iliac artery (eia).A 40mhzh atlantis¿ 018 imaging catheter was selected to view target lesion.During procedure, a non bsc guide wire was advanced to the target lesion.Then the atlantis¿ imaging catheter was used;however it was noted that the device got stuck at the lesion.The physician pushed the imaging catheter and lost image occurred.Dilatation was then performed with an unspecified balloon and the imaging catheter was exchanged to another of the same device.Finally, stent deployment was performed and good blood flow was confirmed.No patient complications were reported and the patient's status is good.
|