Evaluation summary: as received, the barrel end of the sizer was completely detached from the rod.The barrel end had two pieces broken, both were 10mm in length.The pieces matched up to the barrel, and there did not appear to be any missing pieces.The replica end had 3 crazes seen at the rod to sizer junction.Additional manufacturer narrative: customer report of broken sizer was confirmed.There is no serial number of lot number available for this device; therefore, the device history record (dhr) review cannot be done.Although these devices are reusable, they do not have an indefinite shelf life.According to the directions in the instructions for use (ifu) supplied with the accessories, sizers and sizer handles must be disassembled before resterilization.The accessories should be examined for signs of wear, such as dullness, cracking, or crazing, and do not use if deterioration is observed.The ifu was reviewed and no inadequacies were identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.No corrective action is applicable to this case; however, edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.No corrective or preventative actions are required.
|