STRYKER ENDOSCOPY-SAN JOSE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080617
Device Problems Crack (1135); Peeled/Delaminated (1454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that insulation at tip of the shaft was cracked / peeling.
 
Manufacturer Narrative
(b)(4).The device manufacture date is not known at this time.Alleged failure: insulation damage.Confirmed failure: cracked/peeling insulation at middle of shaft.Probable root cause: poor autoclave reliability.Incorrect sterilization/reprocessing procedure.Handling procedures.Contact forces.Product used beyond defined useful life.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that insulation at tip of the shaft was cracked / peeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM, 33CM PEEK MONOPOLAR HANDLE 33CM
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key5553003
MDR Text Key42542797
Report Number0002936485-2016-00323
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080617
Device Lot Number¿1241405¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/11/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-