BAXTER HEALTHCARE - MOUNTAIN HOME CAPD DISCONNECT Y SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number JIC8328 |
Device Problem
Misconnection (1399)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/16/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a connection issue between a uv flash transfer set and a capd disconnect y set occurred while using a uv flash machine.The two devices were not able to be connected properly and when the uv flash machine was opened, a misspike was found with a bent spike.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed with naked eye and magnification with no issues noted.Functional testing was performed including leak testing, clear passage test, clamp function test and set connection and disconnection.All functional testing performed was acceptable.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|