Catalog Number JIC8328
Device Problem Misconnection (1399)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/16/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a connection issue between a uv flash transfer set and a capd disconnect y set occurred while using a uv flash machine.The two devices were not able to be connected properly and when the uv flash machine was opened, a misspike was found with a bent spike.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed with naked eye and magnification with no issues noted.Functional testing was performed including leak testing, clear passage test, clamp function test and set connection and disconnection.All functional testing performed was acceptable.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPD DISCONNECT Y SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5562757
MDR Text Key42159015
Report Number1416980-2016-06792
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberJIC8328
Device Lot NumberH15I09070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received05/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-