As reported by our (b)(4) affiliate, during deployment of a 23mm sapien 3 valve, a leak at the handle of the delivery system was noted resulting in partial deployment of the valve.Post dilatation was performed and no paravalvular leak (pvl) was observed.There was no consequence to the patient.
|
The delivery system was returned to edwards lifesciences for evaluation.During visual inspection, a break was noted on the balloon shaft distal to the y-connector.A leak was confirmed under the balloon shaft strain relief at the y-connector, which resulted from a complete break on the balloon shaft distal to the y-connector.In addition, a slight compression was observed on the flex catheter near the flex tip.No other abnormalities observed.The balloon shaft outer diameter (od) distal to the y-connector bond met specification.No manufacturing non-conformances were found which could have contributed to the reported complaint event.No ifu/training deficiencies were identified.A review of complaint history from (b)(6) 2015 to (b)(6) 2016 revealed other returned complaints confirmed for commander delivery systems leakage distal to the y-connector (all sizes).In addition, a review of complaint history for the month of (b)(6) 2016 revealed that the occurrence rate for the trend category of commander delivery system leakage exceeded the trend category control limits.All devices are 100% inspected by manufacturing and quality from the distal to proximal ends of the device under 2.85x minimum magnification for device damage (e.G.Kinks, cuts).Also, balloon catheters are 100% leak tested.The related manufacturing lot underwent product verification testing on a sampling plan, which included visual inspection, tensile testing of the y-connector/balloon shaft joint, and inflation and deflation time; all samples passed visual inspection.During tensile testing, the calculated tolerance limit was statistically above the lower specification limit.The complaint for delivery system leakage was confirmed as a break in the balloon shaft adjacent to the y-connector was observed.Although it is possible that the balloon shaft was bent, kinked, or pulled during the procedure to cause or propagate the observed damage, it is also possible that a potential manufacturing/design related issue may have also contributed to the complaint event.Due to the severity associated with balloon shaft fracture near the y-connector, and other similar confirmed complaints, a risk assessment was performed and corrective/preventative actions are being addressed through a capa.This unit was manufactured prior to the implementation of any changes.
|