|
Model Number B1371 |
Device Problem
Blocked Connection (2888)
|
Patient Problem
Patient Problem/Medical Problem (2688)
|
Event Date 01/30/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Visual analysis: (b)(4) 2016 - received two (1) - one used one new b1371, 8" ext set w/0.2 micron filter, luer lock, non-dehp tubing, lot# unknown.No mating devices were returned.Blood and a clot was observed in the used filter.The filter was flushed without issue.Functiona'/perfomrance testing: the used set the blood/blood clots were flushed out and decontaminated.The set was then primed with water using gravity pressure (1.5 psig).And then leak tested with water at 25 psig for two (2) minutes.No leaks were found at the filter set.The packaged (unused) set was primed with water using gravity pressure (1.5 psig).The set was then leak tested with water at 25 psig for two (2) minutes.No leaks were found at the filter set.The two (2) devices were tested per the product performance specification.Analysis summary: visual analysis of the "as received" used b1371 complaint sample confirmed blood/blood clots were present in the sets filter components.Following decontamination/flushing the returned sample was tested by engineering to the applicable product specifications and the device performed as intended.With the device performing as intended, the only way for the blood to back flow into the filter is if there was a deliberate disconnect in the line.
|
|
Event Description
|
Complaint received regarding one b1371, 8" ext set w/0.2 micron filter, luer lock, non-dehp tubing, lot# unknown.The report states, "client on hpt program.Client noted blood in proximal tubing and filter of picc line.Previous occurrence this week requiring filter change by rn.Client came back to clinic with obvious clotting present.Unable to aspirate or flush picc line.Client sent to er for tpa instillation for occluded picc.No serious adverse consequences reported.
|
|
Search Alerts/Recalls
|
|
|