Catalog Number EZ10G03
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/23/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Conclusion: to date, the device has not been returned.If the product is returned for evaluation, any further info derived from the evaluation will be submitted in a supplemental 3500a form.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Event Description
It was reported that the patient underwent an unknown procedure on (b)(6) 2016 and suture was used.During the procedure, there were tiny pinholes in the packaging.The procedure was completed with another like device.There were no patient consequences reported.
 
Manufacturer Narrative
The actual device was returned for evaluation.Visual examination observed small holes in the opened foil package.The holes were located in areas where the actual foil presented a multitude of wrinkles likely caused by foil opening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOLOOP LIGATURE WITH COATED VICRYL
Type of Device
LAPROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente dutra, km 1
54
sao paolo 12240 -908
BR   12240-908
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key5569617
MDR Text Key42438266
Report Number2210968-2016-07953
Device Sequence Number1
Product Code GEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K925914
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Catalogue NumberEZ10G03
Device Lot NumberAC2756
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-