• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC. Back to Search Results
Catalog Number 26-1221
Device Problems Use of Device Problem (1670); Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 03/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4) upon completion of the investigation a follow up report will be filed.
 
Event Description
The device did not disengage properly during use.However, there were no adverse consequences to a patient.The device in question was discarded at the facility, and two other perforators from the same lot as this device will be returned for evaluation.
 
Manufacturer Narrative
(b)(4).Upon completion of the investigation it was noted that the customer¿s complaint of "the device did not disengage properly during use" was not verified.This perforator met functional test acceptance requirements; proper engagement and disengagement was achieved with every drill hole and there was no erratic and poor cutting action.The device history records for the perforator was reviewed and all test and inspections associated with the assembly process met specification requirements.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key5573428
MDR Text Key42525865
Report Number1226348-2016-10291
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberLH005S (2)
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-