(b)(4).The manufacturer requested additional information on the reported event, including the product details.Unfortunately no information could be obtained so far.Based on the information available to the manufacturer at this time the health care professional at the hospital noticed the reported battery capacity problem during the set-up of the device and still decided to use the device, supported by an external battery supply.Based on this information the user used the device outside the intended use specified in the instructions provided by the manufacturer (ifu), in modifying the system by connecting an external portable battery supply which is not part of the approved system.The following instructions are provided within the ifu: "only attach the intended components to the rotaflow console.To ensure patient safety, only use tested and approved devices, parts, accessories, and disposables.Take damaged devices out of service immediately and have them tested by the authorized service personnel.The operator is not permitted to change or modify the device.Do not connect equipment which does not form part of this system." a supplemental report will be provided if additional information becomes available.
|
It was reported that while setting up the rotaflow the user noticed upon unplugging the unit, the battery would not hold power and the unit shut off.The account plugged the unit into a portable battery which allowed them to transport the patient within the hospital.The unit, is in use and continues to be in use.(b)(4).
|