• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR FLOTRAC SENSOR; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR FLOTRAC SENSOR; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR Back to Search Results
Model Number MHD8
Device Problems Particulates (1451); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/24/2016
Event Type  malfunction  
Manufacturer Narrative
The lot number was not provided; therefore, a review of the manufacturing records could not be completed.The device evaluation is anticipated.However the complaint could not be confirmed without the completion of the product evaluation.A supplemental report will be forthcoming with the evaluation results.
 
Event Description
It was reported that a white rubber-like material was found inside the fluid path near the pull tab before use.There were no patient complications reported.
 
Manufacturer Narrative
The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Received one flotrac sensor with attached iv tubing set which was coiled with attached paper band.The white particulate was found inside of the female luer of flotrac housing.The kit was continuously flushed for five minutes.The particulate moved underneath the poppet of flush device but still stayed within the flotrac housing after flushing.Per chemistry study (b)(4) the ir spectrum of the unknown white particulate showed similar absorption characteristics when comparing to polyvinyl chloride (pvc) like material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOTRAC SENSOR
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR  
Manufacturer Contact
brian hurd-servin
1 edwards way
irvine, CA 92614
9492506423
MDR Report Key5577404
MDR Text Key43329568
Report Number2015691-2016-01209
Device Sequence Number1
Product Code DRS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K043065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMHD8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-