Model Number PI-RET
Device Problems Suction Problem (2170); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The patient interface (pi) suction ring may lose suction during a procedure.Label copy states corneal fixation vacuum loss can occur.There are several factors that may contribute to suction issues such as doctor¿s technique in applying the suction ring to the cornea, doctor¿s technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient¿s cornea and the suction ring.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
The account reported galvo error and suction loss during surgery with an intralase patient interface (pi) after the patient was applanated and after the laser fired.It is also noted that the procedure was completed successfully.There was a flap/ring replaced.There was no patient injury reported and no surgical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRALASE
Type of Device
PATIENT INTERFACE
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
valerie sedzicki
1700 east st. andrew place
santa ana, CA 92705
7142478567
MDR Report Key5579140
MDR Text Key42766078
Report Number2648035-2016-00594
Device Sequence Number1
Product Code HNO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/08/2017
Device Model NumberPI-RET
Device Lot NumberCA14718
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-