DEPUY SYNTHES POWER TOOLS 11 CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number QD11
Device Problems Overheating of Device (1437); Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Date 11/19/2014
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to device worn from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during service and repair, it was observed that the 11 cm angle attachment had bearings defective and ball bearings heating up.This event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11 CM ANGLE ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5581984
MDR Text Key43560430
Report Number1045834-2016-11309
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
PK011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 11/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberQD11
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received11/19/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-