SYNTHES OBERDORF RECIPROCATING SAW ATCH FOR PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.040
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to normal wear.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the reciprocating saw attachment device was frozen and not functioning.During an in-house engineering evaluation, it was observed that the device was found to run but heated up.It was also noted that the device had damaged bearings and worn and damaged drive shaft and eccentric coupling.It was also noted that the device had a damaged needle bearing (one needle came out) and worn cone sleeve.The event was not reported to have occurred during surgery.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECIPROCATING SAW ATCH FOR PEN DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5582723
MDR Text Key43633042
Report Number8030965-2016-13036
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PK043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.040
Device Lot Number1384121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2016
Date Manufacturer Received03/31/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-