Product event summary: the electrophysiology (ep) catheter and data files were returned and analyzed.The data files confirmed the system notice that indicated that the injection was stopped because the cryomapping temperature exceeded the preset value by more than fifteen degrees (b)(4) and the system notice indicating that the safety system detected a high level of refrigerant flow and stopped the injection (b)(4).Visual inspection of the catheter showed that the shaft of the catheter was broken 1.2 inches from the strain relief.The catheter failed the performance test due to a broken/snapped shaft.There were no traces of blood, and no patient complications.Also, the shaft material was brittle and uv sensitive when exposed for a long period of time to the light.In conclusion, the reported issue was confirmed through testing and data analysis.The catheter failed inspection due to a shaft breakage/snapping.
|
It was reported that during a cryo ablation procedure, a system notice had occurred indicating that the injection was stopped because the cryomapping temperature exceeded the preset value by more than fifteen degrees.Also, there was temperature issues and an indicating that the safety system detected a high level of refrigerant flow and stopped the injection.The electrophysiology (ep) catheter was replaced and the procedure was completed with the console.The device was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|