Model Number ADVIA CHEMISTRY LIPASE REAGENTS |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/23/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cause of the falsely high lipase result on the one patient sample is unknown.Siemens healthcare is investigating the issue.
|
|
Event Description
|
The customer has obtained a falsely high result on one patient sample for lipase assay on the advia chemistry 1800 instrument when using reagent lot 354459.The sample was run in a panel with other lipid tests when the falsely high result was obtained the sample was repeated and run on its own and the value was lower.The initial high result was not reported to the physician(s).It is unknown if the repeat value was reported to the physician(s).There were no reports of patient intervention or adverse health consequences due to the falsely high result for lipase.
|
|
Manufacturer Narrative
|
Additional information (4/28/2016): a siemens headquarters support center (hsc) representative evaluated the instrument files and confirmed that the contamination avoidance and detergent settings on the instrument were correct per assay instructions.A siemens field service engineer (fse) visited the customer site and evaluated the instrument.The fse changed the dilution tray, reaction tray cuvettes, inspected reagent 1 probe and reagent 2 probe and mixer paddles.Reagent 1 probe was replaced by the fse.The customer has notified siemens that they will be using the lipase assay on its own and will not be running it with other assays.The cause of the falsely high result for lipase is unknown.
|
|
Search Alerts/Recalls
|