It was reported that the systolic arterial pressure value shown on the nihon kohden patient monitor was below 50mmhg, while the expected value should have been over 100mmhg during use.The value and the waveform both matched and the waveform was displayed as overdamping.It is unknown if there were any error messages observed.There was no problem zeroing before use and no occlusion, kinks or leaks were observed on the line.It was further indicated that the clinician was going to remove the needle to make another insertion site, but the problem was resolved by exchanging the flotrac unit and the cable.The sample of the tracing was not available.The patient was not treated based on the inaccurate value and there were no patient complications reported.
|