Model Number INFKIT2
Device Problems Defective Alarm (1014); Failure to Infuse (2340)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2016
Event Type  malfunction  
Manufacturer Narrative
The pump was returned to mmdg for evaluation, and subjected to visual, mechanical, and flow-related testing.The device's dhr and internal event log were also reviewed.The pump passed all tests, and was found to be performing according to specifications.No errors pertaining to the event as described by the initial reporter appeared in the event log, and the device's dhr showed no nonconformances.Mmdg is unable to replicate the complaint.
 
Event Description
The initial reporter stated that the pump acted as if it was running, but the bag was still full.The pump alarmed dose done as expected, but nothing was actually dispensed.No further information was provided.No indication of patient harm.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ENTERAL FEEDING PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
matt brinkerhoff
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key5592849
MDR Text Key43204556
Report Number1722139-2016-00417
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 04/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/23/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-