• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND TELIGEN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND TELIGEN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number F102
Device Problems Device Alarm System (1012); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/06/2016
Event Type  Injury  
Manufacturer Narrative
The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
 
Event Description
Boston scientific received information that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device was explanted and will be returned for analysis.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to a product performance anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This particular device was included in the advisory population.
 
Manufacturer Narrative
This device was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Review of the device memory indicated that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.A series of automated electrical/functional tests were conducted and no issues with device performance were observed; basic sensing, pacing and shocking functions of the device were verified.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.
 
Event Description
--.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELIGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5596191
MDR Text Key43276633
Report Number2124215-2016-01135
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/17/2010
Device Model NumberF102
Other Device ID NumberTELIGEN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received02/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0026-2014
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-