Model Number 0431157-51 |
Device Problem
Retraction Problem (1536)
|
Patient Problem
Awareness during Anaesthesia (1707)
|
Event Date 04/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Event took place in (b)(6) and has been reported through (b)(4) distribution subsidiary pajunk medical produkte (b)(4).Currently the data is poor and the device has not been sent back/ analysed.As soon as further data will be available a follow up report will be sent into the agency.
|
|
Event Description
|
(b)(4).Tentative summarizing translation of initial reporter's narrative: ülacement of catheter for continuous epidural anaesthesia, unable to draw back needle over catheter; stylet got hooked in inner coil.Catheter had to be replaced, procedure re-initialized.
|
|
Manufacturer Narrative
|
With one of the devices involved an individual device failure can be confirmed.The second catheter has been stretched and ruptured due to handling-based complications.Based on risk assessment and clinical evaluation file is considered as closed.
|
|
Event Description
|
(b)(4).Tentative summarizing translation of initial reporter's narrative: placement of catheter for continuous epidural anaesthesia, unable to draw back needle over catheter; stylet got hooked in inner coil.Catheter had to be replaced, procedure re-initialized.Amendment: 2 catheter kits involved.Lot is unknown, packaging has been disposed.
|
|