Model Number ACUSON SEQUOIA S512
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2014
Event Type  malfunction  
Manufacturer Narrative
This failure was traced to a failed component (shorted capacitor) in the monitor.
 
Event Description
There is a burning smell coming from the system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSON
Type of Device
DIAGNOSTIC ULTRASOUND DEVICE
Manufacturer (Section D)
SIEMENS MEDICAL SOLUTIONS USA INC.
mountain view CA 94043
Manufacturer Contact
karen smith, vp
685 east middlefield road
mountain view, CA 94043
8004228766
MDR Report Key5607603
MDR Text Key43767856
Report Number3009498591-2015-00150
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
052331
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Not Applicable
Remedial Action Replace
Type of Report Initial
Report Date 09/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACUSON SEQUOIA S512
Device Catalogue Number08245875
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2004
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-