Results: the suction connector to the penumbra system max aspiration tubing (tubing) was blocked.Conclusions: evaluation of the returned device revealed that the tubing suction connector was blocked.The tubing was connected to the aspiration pump and no vacuum was present.Further evaluation revealed that the suction connector to the tubing was blocked by the adhesive that was used to bond the connector to the tubing.The blocked suction connector likely prevented the vacuum pressure from being present during preparation.No further damage was observed on the tubing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a thrombectomy procedure, the physician noticed that the penumbra system max aspiration tubing (tubing) had a hole and therefore, suction was not possible.The damaged tubing was found prior to use and was not used for the procedure.The procedure was completed using a new aspiration tubing.
|