• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number MICROBLENDER
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Date 04/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).At this time, carefusion has not received the device from the customer.In the event the device is received for evaluation or additional information is received a follow-up report will be submitted.
 
Event Description
The customer stated that while testing the unit it was found that the device does not have an audible alarm when it should.There was no patient involvement at the time of the incident.
 
Manufacturer Narrative
Results of investigation: the factory service department received the blender for evaluation and found a damaged alarm reed plate.The reed plate was routed to the failure analysis lab however due to the damage an investigation was unable to be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
7149227830
MDR Report Key5626483
MDR Text Key44730824
Report Number2021710-2016-03668
Device Sequence Number1
Product Code BZR
Combination Product (y/n)N
PMA/PMN Number
K883038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMICROBLENDER
Device Catalogue Number10041A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-