CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Defective Device (2588); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that patient's vns generator was unable to be interrogated using two different programming systems.It was reported that one of the programming systems showed the error message: "unable to interrogate with this version of software - error code 310".Troubleshooting was performed on both programming system by the company representative.The programming tablet usb cable was faulty and was replaced.The programming wand 9 volt battery was changed after what the wand was confirmed to be working.It was reported that the patient will be seen again in clinic to interrogate the generator.No generator issue is suspected.Return of the usb cable for analysis is expected, but it has not been received to date.
 
Event Description
The suspected usb cable was received by the manufacturer.An analysis was performed on the returned usb to db9 cable and the reported allegation was verified.The cause for the reported allegation is associated with two disconnected wire connections in the returned serial cable.Once the wires were soldered onto the serial cable pcb, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5629884
MDR Text Key45288880
Report Number1644487-2016-00946
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-