MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701028718
Device Problems Insufficient Heating (1287); Device Maintenance Issue (1379)
Patient Problem No Patient Involvement (2645)
Event Date 05/02/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A maquet field service technician was on site and investigated the unit.The technician troubleshot the device and found a the valves polluted.The valves were cleaned and tested for functionality, the main circuit heated up like it should.The device was tested for electrical safety and the temperatures were checked.All tests were performed successfully.A supplemental medwatch will be submitted if additional informations becomes available.
 
Event Description
It was reported that the main circuit of the hcu30 did not heat.The incident occurred during preventive maintenance so there were no patient involved.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU40 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5633961
MDR Text Key44910452
Report Number8010762-2016-00318
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K031544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701028718
Device Catalogue Number939001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-