ST. JUDE MEDICAL, INC. (CRM-KISTA) FORTIFY DR, U1.6 DF1 ID; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2233-40
Device Problems Premature Discharge of Battery (1057); Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the device could not be interrogated.Device was explanted and replaced.Patient's condition was good.
 
Manufacturer Narrative
No conclusion code available; the reported field event of an inability to interrogate the device and was confirmed in the laboratory.Upon receipt, communication could not be established between the device and programmer due to the battery being depleted.Based on all available parameter and usage information, device longevity was found to be below the expected limits.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY DR, U1.6 DF1 ID
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
ulla strindlund
box 7051
isafjordsgatan 15
kista SE-16-407
SW   SE-16407
4684744043
MDR Report Key5634678
MDR Text Key44511902
Report Number3010215456-2016-03223
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 06/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2012
Device Model NumberCD2233-40
Device Lot Number3290186
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0115-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-