• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHI; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHI; SWAN-GANZ CATHETER Back to Search Results
Model Number 777HF8J
Device Problem Low Readings (2460)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/08/2016
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.5 cc limited volume syringe and three non-edwards plugs were returned for evaluation.An arrow introducer with non-edwards contamination shield was located on the catheter body between 37.5 cm and 91 cm proximal from the catheter tip.No packaging was returned.No fault messages were observed on the lab vigilance ii monitor when the catheter was connected.The thermistor was found to read 36.9 c when submerged into a 37.0 c water bath.The catheter ran cco in 37.0 c water bath on vigilance ii monitor for 5 minutes without error.The thermistor and thermal filament circuit were continuous, there were no open or intermittent conditions.No visible inconsistency was observed on eeprom data.Resistance value of thermal filament circuit was within specification, measuring 39.52 ohms.Both the thermistor and thermal filament connectors were opened and no visible inconsistencies were found.No visible damage to the catheter body, balloon, or returned syringe was observed.All through lumens were patent without any leakage or occlusion.The balloon inflated clear, concentric and remained inflated for more than 5 minutes without leakage.Balloon inflation test was performed using returned syringe with 1.5 cc air.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.The complaint could not be confirmed.No evidence of a manufacturing nonconformance was noted.It could not be determined if procedural factors or device handling may have contributed to the reported event.No actions will be taken at this time.
 
Event Description
It was reported that the swan ganz catheter was placed in the patient after cardiac surgery.After two to three days of use, cci value gradually decreased from about 2.5l/min/m2 to 1.5l/min/m2.The expected cci value was more than 2.5l/min/m2.The catheter position was checked and the catheter was in the correct position.The patient was not treated based on the incorrect value.Error message was not observed.Vig2 monitor and the cable were tested by the customer but there were no problems observed.No further information was able to be obtained.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHI
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key5638601
MDR Text Key45313624
Report Number2015691-2016-01525
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number777HF8J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-