• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97702
Device Problems Premature Discharge of Battery (1057); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2016
Event Type  Injury  
Event Description
Both the patient and healthcare professional (hcp) reported through a manufacturer representative that the patient¿s implantable neurostimulator (ins) had reached end of service (eos) on (b)(6) 2016.It was stated the ins experienced premature battery depletion.It was noted the patient ¿had never had a connection issue and was extremely satisfied by the clinical outcome.¿ it was further noted the patient had utilized one stimulation group, which was set to continuous stimulation.Impedance testing was performed and found impedances were in the range of ¿700-900 ohms.¿ the cause of the premature depletion was unknown.Replacement of the ins was planned in the future as a result of the event, though it was unknown whether the replacement had been scheduled yet at the time of report.The issue remained unresolved at the time of report and the patient was alive with no injury at that time.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information received indicated the product was received by the company at an international site, and was shipped to a different site for analysis; however the product has not been received at the site to perform analysis.The current device location is unknown and an investigation is ongoing to locate the device.
 
Manufacturer Narrative
(b)(4) has been removed.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Device analysis for ins (b)(4) revealed no significant anomaly.The ins battery was at normal end of life.Telemetry and output was okay.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key5641543
MDR Text Key44776772
Report Number3004209178-2016-09134
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2016
Device Model Number97702
Device Catalogue Number97702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-