• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT¿; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT¿; STENT, CAROTID Back to Search Results
Model Number H965SCH647130
Device Problem Device Packaging Compromised (2916)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).
 
Event Description
It was reported that the sterile package was compromised.During the unpacking of a 10.0-31 carotid wallstent, iit was noted that the inner package has an air-leakage condition.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Manufacturer Narrative
Device evaluated by mfr.: an examination of the packaging identified that the tyvek pouch was opened and there was blood visible on the pouch.This blood is consistent with the tray/packaging having been opened in a hospital environment.The opened pouch was also analyzed for punctures, tears or holes with no issues noted.No further issues were identified during product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
 
Event Description
It was reported that the sterile package was compromised.During the unpacking of a 10.0-31 carotid wallstent, it was noted that the inner package has an air-leakage condition.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT¿
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5644426
MDR Text Key44893164
Report Number2134265-2016-03847
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/23/2018
Device Model NumberH965SCH647130
Device Catalogue NumberSCH-64713
Device Lot Number0018128230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-