Model Number H965SCH647130 |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/11/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.(b)(4).
|
|
Event Description
|
It was reported that the sterile package was compromised.During the unpacking of a 10.0-31 carotid wallstent, iit was noted that the inner package has an air-leakage condition.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: an examination of the packaging identified that the tyvek pouch was opened and there was blood visible on the pouch.This blood is consistent with the tray/packaging having been opened in a hospital environment.The opened pouch was also analyzed for punctures, tears or holes with no issues noted.No further issues were identified during product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
It was reported that the sterile package was compromised.During the unpacking of a 10.0-31 carotid wallstent, it was noted that the inner package has an air-leakage condition.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
|