DEPUY SYNTHES POWER TOOLS ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Temperature Problem (3022)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition that the device was locking up/freezing was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device was heating up during pre-test.It was further noted that the device marking/labeling was faded, and the device failed a hand switch test.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that during pre-surgery testing, it was observed that the handpiece device was locking/freezing up.During in-house engineering evaluation, it was found that the device was heating up.It was reported that there were no delays in a surgical procedure and an unspecified spare device was available.There was no patient involvement.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|