Catalog Number UNK-HIP |
Device Problems
Fracture (1260); Loss of Osseointegration (2408); Osseointegration Problem (3003)
|
Patient Problems
Pain (1994); Inadequate Osseointegration (2646); No Code Available (3191)
|
Event Date 06/10/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is the subject of litigation or a legal claim and currently complete product detail is not available at this time.A follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Litigation received.Litigation alleges femoral stem fracture and pain.
|
|
Manufacturer Narrative
|
Conclusion and justification status for mdr: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The investigation could not draw any conclusions regarding the reported event.Based on the inability to determine a root cause, the need for corrective action was not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).
|
|
Event Description
|
Ppf alleges component fracture, loosening of cup, and loosening of stem.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: h6 (patient).Patient code: no code available (3191) was used to capture device revision or replacement.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: investigative update: (b)(6) 2021.An image of explanted device(s) has been provided for review.No device associated with this report was received for examination.A photo of explanted devices has been provided for review.This includes the femoral stem which is fractured, as reported.It was not possible to identify any product / lot information.The stem is fractured just proximal of mid-shaft.It is not possible to conclusively determine a root cause for the fracture using photographic images.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a device history record (dhr) review, was not possible because the required lot code(s) was not provided.
|
|
Search Alerts/Recalls
|