PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DEFIBRILLATORS, AUTOMATIC, EXTERNAL Back to Search Results
Model Number 20
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 04/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's therapy connector assembly and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio further examined the removed therapy connector assembly and confirmed that a broken pin was stuck in the connector.As a result, another quik-combo therapy cable could not be connected to the device.It was later confirmed by the customer, a biomedical engineer, that the damaged quik-combo therapy cable was disposed of on-site and therefore not available for further examination by physio-control.
 
Event Description
The customer contacted physio-control to report that a pin from a quik-combo therapy cable had broken off of the cable and become lodged in their device's therapy connector assembly.As a result, defibrillation may not be possible.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATORS, AUTOMATIC, EXTERNAL
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 9706
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 9706
Manufacturer Contact
yuchol kim
4258674000
MDR Report Key5648571
MDR Text Key45041980
Report Number3015876-2016-00549
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2016
Is the Reporter a Health Professional? Yes
Device Age11 YR
Event Location Hospital
Date Manufacturer Received04/19/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-