Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/11/2016 |
Event Type
No Answer Provided
|
Manufacturer Narrative
|
The facility stated that water was backing up within the lubricant bottle.No water was reported outside of the bottle.A steris service technician arrived on site, inspected the unit and identified that water pressure was above the specification for the unit causing the reported event.The technician informed the customer that incoming water exceeded steris water pressure specification for the caviwave pro ultrasonic unit.The unit was removed from service until a pressure regulator can be installed by the user facility's maintenance department.The unit is not under steris contract for maintenance services and all service is performed by the user facility.No additional issues have been reported with the unit.
|
|
Event Description
|
The user facility reported water was backing up in their caviwave pro ultrasonic.No injury, procedural delay, or cancellation was reported.
|
|