• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS STERNAL SAW SYSTEMS; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL-SARNS STERNAL SAW SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS STERNAL SAW SYSTEMS; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL-SARNS STERNAL SAW SYSTEMS Back to Search Results
Model Number 5590
Device Problems Failure To Service (1563); Maintenance Does Not Comply To Manufacturers Recommendations (2974)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is related to emdr #1828100-2016-00210.The reported complaint was confirmed.The sternal saw drive components were worn due to lack of preventive maintenance (pm) and lubrication.Per the instructions for use (ifu), the sternal saw must have pm every six months for optimum operation.The srt replaced the cam/gear assembly and performed verification / release testing of the sternal saw.The unit operated to manufacturer specifications and was returned to clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The service repair technician (srt) reported that during routine testing of the device at the service center, the cam/gear assembly has faulty bearings of the sternal saw (does not rotate smoothly).There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS STERNAL SAW SYSTEMS
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL-SARNS STERNAL SAW SYSTEMS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5656638
MDR Text Key45881121
Report Number1828100-2016-00347
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K935391
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number5590
Device Catalogue Number5590
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-